Proceedings Vol. 22 (2016)
ENGINEERING MECHANICS 2016
May 9 – 12, 2016, Svratka, Czech Republic
Copyright © 2016 Institute of Thermomechanics, Academy of Sciences of the Czech Republic, v.v.i., Prague
ISSN 1805-8248 (printed)
ISSN 1805-8256 (electronic)
list of papers scientific commitee
pages 25 - 28, full text
This article demonstrates results of the tests of mechanical properties of steel guidewires used in cardiac surgical procedures. The established research methodology embracing specimen geometry and simulated operating environment allowed determining the maximum strength of a guidewire during surgeries. The results of the tests were subjected to statistical analysis, on the basis of which the differences in mechanical strength of material were determined. Results indicate differences in the strength of cardiac guidewires which are in the same group characteristics and end uses. This study considered on three types of steel guidewire strength: Amplatz Ultra Stiff; Lunderquist Extra Stiff; Back – Up Meier Steerable. The AMPLATZ guidewire showed the lowest tensile strength, stiffness and the highest torsion with Young's modulus. The BACK – UP MEIER guidewire has very similar properties to the LUNDERQUIST guidewire.
back to list of papers
Text and facts may be copied and used freely, but credit should be given to these Proceedings.
All papers were reviewed by members of the scientific committee.